4-20
IM CA71-E
4.7
Sweep Function
The sweep function varies the output in a linear manner. For further
details, see Section 7.1, “Sweep Function.”
4.8
Auto Step Function
The auto step function varies the output in a step-by-step manner. For
further details, see Section 7.2, “Auto Step Function.”
4.9
Temperature Monitor Function
Using the
key, you can show the monitored temperature on the
LCD, as described below.
◆
When the Voltage, Current, Resistance or Pulse (Continuous
Pulse Train or Number of Pulses) Range Is Selected
The reading of a sourced signal remains changed to the temperature
detected by the built-in temperature sensor of the calibrator as long as
the
key is kept held down. Thus, you can monitor the room’s tem-
perature.
◆
When the Temperature (Thermocouple or RTD) Range Is Selected
•
Pressing the
key once allows you to monitor the electromotive
force (mV) or resistance (
Ω
) equivalent to the preset temperature.
The monitored value does not reflect the correction made by the RJ
sensor.
•
Pressing the
key once again changes to the temperature de-
tected by the RJ sensor connected to the calibrator or the internal
temperature of the calibrator.
•
Pressing the
key one more time reverts to the initial normal
setting mode.