IM CA71-E
7-5
Functions Provided by DIP Switch
7
Step 8: Using each pair of
and
keys, set the value of the de-
nominator
m and the starting setpoint of the enumerator n.
(See Section 4.6, "Divided Output Function (
n/m), for further
details.) The starting setpoint is the minimum of the variable
n
for auto step operation.
Step 9: Pressing the
key initiates the automatic stepping of the
divided output (
n/m), as described below.
Assuming the starting value of the variable
n is 1, the output
cyclically changes with the variable
n as n changes in the fol-
lowing manner.
n = 1
→
2
→
3
→
· · · (
m – 1)
→
m
→
(
m – 1)
→
· · ·
→
2
→
1
→
2
→
3
→
· · ·
Step 10: To quit the auto step function, turn off the calibrator by press-
ing the
key. Place switch 1 back in the OFF (left-side)
position. Note that disabling the divided output mode (
n/m)
before placing switch 1 back in the OFF position enters
sweep mode.
TIP
• Pressing the
key to change to the SOURCE OFF state pauses the auto step
operation.
• To execute the auto step function again, press the
key. The auto step opera-
tion resumes from the value of n shown on the LCD.