8-8
IM CA71-E
SD
Sets sourced setpoint/queries the current setting.
When normal
condition
Syntax for setting
SDm<delimiter>
Syntax for query
SD?<delimiter>
⇒
Response:
SDm<delimiter>
Description of parameter
m: Sourced setpoint (7 digits) ex. +1.0000
SF
Queries the source function.
When normal
condition
Syntax for query
SF? <delimiter>
⇒
Response:
SFm<delimiter>
Description of parameter
m: Function
m=0: 30V
m=1: 10V
m=2: 1V
m=3: 100mV
m=4: Resistance
m=5: Pulse
m=6: 20mA
m=7: 4 to 20mA
m=8: 20mASINK
SO
On/Off of source output/queries the current setting.
When normal
condition/adjustment
Syntax for setting
SOm <delimiter>
Syntax for query
SO? <delim iter>
⇒
Response: SOm <delimiter>
Description of parameter
m: Condition of generation
m=0: Off
m=1: On
SR
Sets the sourcing range/queries the current
setting.
When normal
condition/adjustment
Syntax for setting
SRm<delimiter>
Syntax for query
SR?<delimiter>
⇒
Response:
SRm<delimiter>
Description of parameter
m: sourcing range
[100mV] (When normal condition)
m=0: DC 100mV
m=1: TcK
m=2: TcE
m=3: TcJ
m=4: TcT
m=5: TcR
m=6: TcB
m=7: TcS
m=8: TcN
m=9:
TcL
m=10:
TcU
[100mV] (When adjustment)
m=0: DC 100m V
m=1: TcK
[
Ω
] m=0:
400
Ω
m=1:
Pt100
m=2:
JPt
[Freq ]
m=0: 500Hz
m=1: 1kHz
m=2: 10kHz
8.7 Detailed Description of Commands
Содержание HANDY CAL CA51
Страница 1: ...User s Manual Model CA51 71 HANDY CAL Calibrators IM CA71 E 1st Edition Feb 2002 IM CA71 E...
Страница 11: ......
Страница 29: ......
Страница 51: ......
Страница 61: ......
Страница 69: ......
Страница 77: ......
Страница 89: ......
Страница 91: ......
Страница 101: ......
Страница 103: ......