IM CA71-E
5-5
Measurement
5
◆
Replacing the Fuse
The current input protection fuse in the mA/3WIRE terminal is housed
inside the fuse holder (labeled FUSE) on one side panel of the calibra-
tor. To replace the fuse, first remove the fuse holder labeled FUSE by
turning the holder counterclockwise with a flatblade screwdriver. Then,
replace the fuse and insert the fuse holder back in place. Fasten the
fuse holder by turning it clockwise. The replacement fuse is described
below.
Part Number
Rating
A1501EF
125 mA/250 V, fast-acting
5.2 Measuring 300 V AC-range Voltage, DC Voltage, AC Voltage or DC Current