<6. CALORIE / PARAMETER SETTING>
6-10
IM 11T03E01-51E
(3) Display Mode
Select one of the following seven parameters to be shown in the message field. Each press of
the [NO] key cycles through the choices in the order shown below:
1. Physical density [XX.XXXX KG/M3] or [XX.XXXX LB/FT3]
(depends on the setting of CODE 21)
2.
Compensated density [X.XXXX KG/M3 ]or [X.XXXXX LB/FT3]
(depends on the setting of CODE 21)
3. Calorific value
[XXX.XXX MJ/M3] or [XX.XXXX KBTU]
(depends on the setting of CODE 22)
4. Temperature [XXX.X °C] or [XXX.X °F]
(depends on the setting of CODE 23)
5. Pressure [XXX.XX KPA] or [X.XXXX MPA] or [XX.XXX PSI]
(depends on the setting of CODE 20)
6. Current for analog output 1 (%) [XXX.X MA1%]
7. Current for analog output 2 (%) [XXX.X MA 2%]
Show the desired choice and press the [YES] key. That choice is set as a parameter to be
measured in the measurement mode (<MEASURE>). Parameters 1, 2 and 3 are configurable,
while parameters 4 to 7 are read-only.