<7. HYDROGEN PURITY / PARAMETER SETTING>
7-18
IM 11T03E01-51E
CODE 04: Setting of Output-smoothing Constants
[
*SMOTH
]
This setting defines the constants for output smoothing.
This smoothing constants is output-smoothing constants of converter (electrical circuit boards).
Configurable range: 00 to 60. (unit : second)
[ENT]
04
*CODE
*SERVC
XX
*SMOTH
[ENT]
[YES]
F0711.ai
*
1
*
2
*
3
*1:
See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
Figure 7.11
Setting of Output-smoothing Constants
CODE 05: Setting of Contact Outputs Status
[
*CNTCT
]
This setting defines the status of the contact outputs.
Configurable range: 00 to 15
[ENT]
05
*CODE
*SERVC
XX
*CNTCT
[ENT]
[YES]
*
1
*
2
*
3
*1:
See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
F0712.ai
Figure 7.12
Setting of Contact Outputs Status