<5. DENSITY / PARAMETER SETTING>
5-6
IM 11T03E01-51E
Table 5.4
Setting Parameters at Service Level
Code
No.
Item
Display
Data to be Set
(or Conditions)
Remarks
20 Selection of pressure unit
*PRES.U
· kPa : 0
· MPa : 1
· psi : 2
See Figure 5.17 for key operation.
21 Selection of density unit
*DENS.U
· kg/m
3
: 0
· lb/ft
3
: 1
See Figure 5.18 for key operation.
22
Selection of calorific value unit
*CAL.U
· MJ/m
3
: 0
· kBTU/ft
3
: 1
See Figure 5.19 for key operation.
23 Selection of temperature unit
*TEMP.U
· °C : 0
· °F : 1
See Figure 5.20 for key operation.
30 Setting of calendar
Setting of year/month/day
Setting of hour/minute
*Y_M_D
*H_M
00.01.01 to 99.12.31
00.00 to 23.59
See Figure 5.21 for key operation.
For leap years, the setting can
cope with ‘February 29’.
31 Showing/hiding of negative measured
values (-)
*MINUS
· Show : 0
· Hide : 1
See Figure 5.22 for key operation.
40
Indication of calibration coefficients
(read-only)
Zero
Span
*C_K_Z
*C_K_S
SX.XXXX
X.XXXX
See Figure 5.23 for key operation.
S of SX.XXXX means Marks such
as a plus or the minus.
41 Indication of oscillation frequency
(read-only)
Frequency F2
Frequency F4
Frequency ratio
*F2.KHZ
*F4.KHZ
*F2/F4
X.XXXXX
X.XXXXX
X.XXXXX
See Figure 5.24 for key operation.
42 Indication of software version
(read-only)
*REV
X.XX
See Figure 5.25 for key operation.
43 Setting of high-resolution mode
*S_CYC
· Normal :
0
· High-resolution : 1
See Figure 5.26 for key operation.
44 Setting of password
*PASS
0.0.0 to 9.9.9
See Figure 5.27 for key operation.
45 Selecting Battery alarm detection
*BAT
· Non detection : 0
· Detection : 1
See Figure 5.28 for key operation.
50 Selecting the instruments
*MODEL
· Density meter : 0
· Calorie meter : 1
· Hydrogen purity, replacement
meter : 2
See Figure 5.29 for key operation.
82 Setting Detector Constant
See Figure 5.30 for key operation.
5.2
Parameter Setting
[Aborting the Setting Operation]
Press the [MODE] key. Normally, the [MODE] key is used to move from the measurement
(<MEASURE>) mode to a mode at the operation level. It is also used to return to the
measurement mode from other modes.
5.2.1
Parameter Setting at Operation Level
To move to the operation level, press the [MODE] key and then enter the password (XXX).
Password is not selected in first condition. When password is necessary, refer to Section 4.2
“Setting Lists” and selected password what you need.