<11. INSPECTION AND MAINTENANCE>
11-4
IM 11T03E01-51E
11.2.2
Inspecting the Analyzer in a FAIL Status
IN GD402, the state of terminals 18 and 19 is changed from close to open if a FAIL status is
detected by the self-diagnostic function during measurement. In addition, the <FAIL> lamp on
the operation panel comes on and the message field shows an error number. If a FAIL status is
detected, take the corrective action given in Table 11.2.
Table 11.2
Corrective Actions when a FAIL Status Is Detected
Message
Shown
Failure Type
Diagnostic Conditions
Corrective Action
Message via
Communication
Err.01
Sensor oscillation
shutdown
Failure to detect
oscillation frequency from the
detector
Reset the power
supply.
Contact Yokogawa
service personnel.
FAULT
SENSOR
Err.02
Error in sensor’s
oscillation
frequency
Frequency F2 : out of
range from 1000 to 10000Hz
Frequency F4 : out of
range from 4000 to 10000Hz
Reset the power
supply.
Contact Yokogawa
service personnel.
OVER
FREQUENCY
Err.03
Error in sensor’s
temperature
Sensor’s failure to detect
temperature
Contact Yokogawa
service personnel.
FAULT TEMP
Err.04
A/D converter
failure
Failure in A/D converter
Contact Yokogawa
service personnel.
FAULT A/D
Err.05
Memory failure
Erroneous data in
EEPROM/EPROM/RAM
Contact Yokogawa
service personnel.
FAULT
MEMORY