<5. DENSITY / PARAMETER SETTING>
5-25
IM 11T03E01-51E
X
*CAL.P
(Selection of calibration item)
XX
*CAL.T
(Setting of calibration time)
XX
*STAB.T
(Setting of stabilization time)
XX.XX.XX
*Y_M_D
(Setting of calibration starting time [year/ month/ day])
XX.XX
*H_M
(Setting of calibration starting time [hour/ minute])
X
*CYCL.U
(Selection of unit of calibration interval)
XXX
*CYCL.T
(Selection of calibration interval)
[ENT]
*SERVC
[ENT]
[YES]
F0517.ai
*1:
See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*3: The number of digits on LCD varies depending on the unit of the calibration interval (parameter [*CYCL.U]).
If the parameter is set to "0" (hour), the indication is two-digit number.
If the parameter is set to "1" (day), the indication is three-digit number.
15
*CODE
[ENT]
[ENT]
[ENT]
[ENT]
[ENT]
[ENT]
*
1
*
2
*
2
*
2
*
2
*
2
*
2
*
2
*
2
*
3
Figure 5.16
Setting of Data for Automatic/Semi-automatic Calibration