1
1-37
IM 701730-51E
I
2
C Bus Signal Analysis Function
1.7
Error Messages
A message may appear during operation. This section describes the meanings of the
messages and their corrective actions. This section lists only the error messages related
to the I
2
C Bus signal analysis function. There are other error messages related to the
DL1700E and communications. These messages are described in the
DL1700E Series
User’s Manual (IM701730-01E)
and the
DL1700E Series Communication Interface
User’s Manual (IM701730-17E)
.
You can set the messages to be displayed in English or Japanese. For the procedure of
setting the message language, see section 15.1, “Changing the Menu Language/
Message Language and Turning ON/OFF the Click Sound” in the
DL1700E Series
User’s Manual (IM701730-01E)
.
If the corrective action requires servicing, contact your nearest YOKOGAWA dealer for
repairs.
Code
Message
Action
Page
37
Aborted the analysis.
–
1-26
38
Data not detected.
–
1-4, 1-10,
Execute again after changing the settings or
1-25, 1-31
reacquiring the waveform.
730
Pattern is not specified.
Enable Address Pattern or Data Pattern in the
1-29
Search Setup dialog box.
739
Analyzed data does not exist.
Execute the analysis.
1-26, 1-31
Execute the analysis.
870
Cannot be specified.
Increase the number of valid bytes.
1-6, 1-11
Invalid byte or bit.