1-15
IM 701610-61E
1
I
2
C-Bus Signal Analysis Function
Combination of Address Pattern, and Data Pattern
Trigger Condition
Address Pattern:
A4
Data Pattern:
27
Condition:
True
Ignore the Restart Condition:
No
Ignore Unexpected Start/Stop Condition:
No
Position:
First Byte
S
27
FE
98
99
27
P
A4
3. Trigger activated here
1. Compare against the specified address pattern (A4). A match
2. Compare against the specified pattern (27). A match
A R/W bit
Combination of Address Pattern, Data Pattern, and Position = Anywhere
Trigger Condition
Address Pattern:
A4
Data Pattern:
99
Condition:
True
Ignore the Restart Condition:
No
Ignore Unexpected Start/Stop Condition:
No
Position:
Anywhere
S
25
AE
57
27
FE
98
99
27
P
A4
3. Trigger activated here
1. Compare against the specified address pattern (A4). A match
2. Compare until a match occurs against
the specified data pattern (99)
A R/W bit
Whether to Ignore the Restart Condition
Trigger Condition
Address Pattern:
F4
Data Pattern:
F5
Condition:
True
Ignore Unexpected Start/Stop Condition:
No
• When Ignore the Restart Condition = Yes
S
A2
Sr
F5
30
27
E4
24
99
55
F4
1. Compare against the address pattern (F4). A match
2. Ignore “Sr.” Compare against the data
pattern (F5). A match
3. Trigger activated here
• When Ignore the Restart Condition = No
S
A2
Sr
F5
30
27
E4
24
99
55
F4
1. Compare against the address pattern (F4). A match
2. Restart trigger detection. Compare against
the address pattern (F4)
1.4 Trigger Setting Examples