1-14
IM 701610-61E
Address Trigger
Trigger Only on the Address Pattern
Trigger Condition
Address Pattern:
A4
Data Pattern:
Not applicable
Condition:
True
Ignore the Restart Condition:
No
Ignore Unexpected Start/Stop Condition:
No
S
25
AE
57
27
FE
98
99
27
P
A4
A R/W bit
Match against the specified address pattern.
Trigger activated here.
Trigger Only on the Data Pattern
Trigger Condition
Address Pattern:
Not applicable
Data Pattern:
27
Condition:
True
Ignore the Restart Condition:
No
Ignore Unexpected Start/Stop Condition:
No
• Position = Anywhere
A R/W bit
S
25
AE
57
27
FE
98
99
27
P
A4
2. Trigger activated here
1. Compare until a match occurs against the specified pattern (27)
• Position = First Byte
S
27
AE
57
27
FE
98
99
27
P
A4
2. Trigger activated here
1. Compare against the specified pattern (27). A match
A R/W bit
1.4 Trigger Setting Examples