3-6
IM CA700-02EN
Setting the 0% Value and 100% Value
For the pressure transmitter output that the CA700 will measure, select the range to calibrate.
11.
Use the ▲ or ▼ key to select
0% Value
, and press
ENTER
.
12.
Press the or
key to move to the digit you want to change, press the ▲ or ▼ key to set the
number.
To confirm the value, press
ENTER
.
Likewise, set the 100% value.
ENTER
ENTER
Setting the Tolerance for Pass/Fail Judgment
13.
Use the
▲
or
▼
key to select
Tolerance
, and press
ENTER
.
14.
Press the or
key to move to the digit you want to change, press the ▲ or ▼ key to set the
number.
To confirm the value, press
ENTER
.
The values assigned to 0% and 100% will be used as references.
Setting Other Item or Confirming
15.
To set items in addition to measurement conditions, press ▲ or ▼ key repeatedly until the
screen changes.
To finish changing the settings, press
Setting Done (F5)
. The settings are confirmed.
3.2 Setting the Device Information and the Calibration Conditions of the Pressure Transmitter