6.6 Status Byte Format
Status byte format (see the explanation of the ESC S command)
bit7
bit6
bit5
bit4
bit3
bit2
bit1
bit0
0
(fixed)
1
(fixed)
Output
error
LOOP
POWER
supply
error
Overrange
Syntax
error
Output
adjustment
completion
Measurement
end
bit7:
Fixed at 0
bit6:
Fixed at
bit:
Becomes if an error occurs during output
The data is held until the status byte is read.
bit4:
Becomes 1 if an error occurs in the power supply for LOOP POWER measurement
The data is held until the status byte is read.
bit:
Becomes if a measured value is outside of its limits (overrange)
Becomes if the CA0 switches to a higher range when auto range is enabled
The data is held until the status byte is read.
bit2:
Becomes when a forbidden operation or command is processed, when the CA0
cannot parse a command, or when a parameter is outside of its range.
The data is held until the status byte is read.
bit:
Becomes when output is on after the output value has changed and the output has
stabilized.
The data is held until the status byte is read.
bit0:
Becomes 1 when measured data is confirmed during measurement.
The data is held until the status byte is read.