4) Press the Hz key to confirm the value.
Press the calibration confirmation key after the reference device’s value stabilizes.
) To calibrate other ranges, press the RANGE key to switch the range, check the function
switch and the input terminals, refer to table , and perform steps and again.
To perform peak hold and loop power calibration, see section 5.1.1, “Peak Hold
Calibration,” and section 5.1.2, “LOOP POWER calibration.”
6) To end calibration, turn the function switch off.
Note
You cannot calibrate the continuity check. Even in calibration mode, normal
measurement is performed. (Keys other than RANGE and SELECT cannot be used.)
Table Calibration Table (Measurements)
Function
Range
Calibration Point (Input value)
AC voltage measurement
6.000 V
6.000 V 0.00/60.00 Hz
60.00 V
60.00 V 0.00/60.00 Hz
600.0 V
600.0 V 0.00/60.00 Hz
000 V
000 V 0.00/60.00 Hz*
DC voltage measurement
6.000 V
6.000 V
60.00 V
60.00 V
600.0 V
600.0 V
000 V
000 V
DC voltage measurement
600 mV
600.0 mV
AC voltage measurement
600 mV
600.0 mV
Diode test
2 V
2.000 V
Resistance measurement
600.0 Ω
600.0 Ω
6.000 kΩ
6.000 kΩ
60.00 kΩ
60.00 kΩ
600.0 kΩ
600.0 kΩ
6.000 MΩ
6.000 MΩ
60.00 MΩ
40.00 MΩ
DC voltage measurement
0.000 mADC
0.000 mA
00.00 mADC
00.00 mA
* If the power supply frequency is 0 Hz, calibrate using 60.00 Hz. If the power supply
frequency is 60 Hz, calibrate using 0.00 Hz.