68
SD
Sets or queries the output value.
Normal Cal.
Command = SDm<CRLF> -> Answer = SDm<CRLF>
Command = SD?<CRLF> -> Return = SDm<CRLF>
Parameters
m (output value) = 0.000 to 25.000
If the CA450 is in a measurement mode, ERR13 is returned.
Î
SF
Queries the output function
Normal Cal.
Command = SF?<CRLF> -> Return = SFm<CRLF>
Parameters
m = 14: Constant current output
15: Current sweep output
If the CA450 is in a measurement mode, ERR13 is returned.
Î
SI
Sets or queries the input voltage value for SENSOR mode
Normal Cal.
Command = SIm<CRLF> -> Answer = SIm<CRLF>
Command = SI?<CRLF> -> Return = SIm<CRLF>
Parameters
m (input voltage value) = 0000 to 9999 (4 digits)
If the CA450 is in an output mode, ERR13 is returned.
Because the decimal place of the input voltage value is fixed,
only the number is entered.
Î
SL
Sets or queries the decimal place for SENSOR mode
Normal Cal.
Command = SLm<CRLF> -> Answer = SLm<CRLF>
Command = SL?<CRLF> -> Return = SLm<CRLF>
Parameters
m (decimal place) = 0: X.XXX, 1: XX.XX, 2: XXX.X, 3: XXXX
If the CA450 is in an output mode, ERR13 is returned.