6
OD
Outputs measured values
Normal Cal.
Command = OD<CRLF> -> Return = abbcsddddddeee<CRLF>
Parameters
Header (4 bytes)*
a = V: voltage, A: current, O: resistance, T: temperature,
F: frequency, S: SENSOR mode
bb = DC: direct current, AC: alternating current,
R3: resistance measurement
c = N: normal, O: overrange, E: no data, B: burnout
* The header will not appear if you specify 0 (no header) in
the H command.
Data Area (10 bytes)
s (sign) = _: plus or –: minus (_ represents a space)
dddddd = measured value (6 digits)
eee = E+0, E+3, E–3, E+6, E–6
(For overrange, no data, and burnout, ddddddeee =
_99999.E+6)
OE
Outputs error data
Normal Cal.
Command = OE<CRLF> -> Return = ERRm<CRLF>
Outputs the most recent output error.
After a value is returned, the saved error number is overwritten
with ‘ERR00<CRLF>’.
Even if there is no error, ‘ERR00<CRLF>’ is returned.
Parameters
m = Error code
See section 6.4, “List of Errors.”