70
SS
Sets or queries the slow step time
Normal Cal.
Command = SSm<CRLF> -> Answer = SSm<CRLF>
Command = SS?<CRLF> -> Return = SSm<CRLF>
Parameters
m = 0: 15 seconds (default)
1: 30 seconds
2: 45 seconds
3: 60 seconds
If you select slow step mode for current sweep output, this
command sets or queries the hold time of each step.
If the CA450 is in an output mode, ERR13 is returned.
Î
ST
Sets or queries the converted value for SENSOR mode
Normal Cal.
Command = STm<CRLF> -> Answer = STm<CRLF>
Command = ST?<CRLF> -> Return = STm<CRLF>
Parameters
m (converted value) = 0000 to 9999 (4 digits)
If the CA450 is in an output mode, ERR13 is returned.
Î
SU
Sets or queries the unit for SENSOR mode
Normal Cal.
Command = SUm<CRLF> -> Answer = SUm<CRLF>
Command = SU?<CRLF> -> Return = SUm<CRLF>
Parameters
m (unit) = 0: A, 1: mA, 2: uA, 3: °C, 5: MΩ, 6: kΩ,
7: Ω, 8: kHz, 9: Hz, 10: uF, 11: nF, 12: %
13: lx, 14: none, 15: V, 16: mV
If the CA450 is in an output mode, ERR13 is returned.