:ACQuire:PWIDth
Function Sets the pulse width or queries the current setting.
Syntax
:ACQuire:PWIDth {<NRf>|AUTO}
:ACQuire:PWIDth?
<NRf> = 3ns to 20us (3E-9 to 20E-6)
Example
:ACQuire:PWIDth AUTO
:ACQuire:PWIDth 3
:ACQuire:PWIDth 3E-9
:ACQuire:PWIDth 3ns
:ACQuire:PWIDth 10us
:ACQuire:PWIDth 10E-6
:ACQuire:PWIDth? ->
:ACQ:PWID 20.0E-06
Description The value that you can specify varies
depending on the wavelength and distance range.
You cannot set the pulse width if the distance
range is set to auto.
The value that you can specify varies depending
on the wavelength and distance range.
You cannot set the pulse width if the distance
range is set to auto.
Auto: AUTO
3 ns:
3E-09
10 ns:
10E-09
20 ns:
20E-09
50 ns:
50E-09
100 ns:
100E-09
200 ns:
00E-09
500 ns:
500E-09
1 us:
1E-06
2 us:
2E-06
5 us:
5E-06
10 us:
10E-06
20 us:
20E-06
:ACQuire:REALtime:STARt
Function Executes the realtime measurement.
Syntax
:ACQuire:REALtime:STARt
Example
:ACQuire:REALtime:STARt
Description This command is valid while the measurement is
stopped.
:ACQuire:REALtime:STOP
Function Stops the realtime measurement.
Syntax
:ACQuire:REALtime:STOP
Example
:ACQuire:REALtime:STOP
Description This command is valid while the realtime
measurement is only in progress.
:ACQuire:SMPinterval:DATA
Function Sets the sampling interval or queries the current
setting.
Syntax
:ACQuire:SMPinterval
:DATA {<NRf>|NORMAL|HI}
:ACQuire:SMPinterval:DATA?
Example
:ACQuire:SMPinterval:DATA 2.0
:ACQuire:SMPinterval:DATA NORMAL
:ACQuire:SMPinterval:DATA? ->
:ACQ:SMP:DATA HI
Description You only set NORMAL or HI if the distance range
is set to auto.
You only set NORMAL or HI if the distance range
is set to auto.
Normal:
NORMAL
High resolution: HI
5 cm:
0.05
10 cm:
0.10
20 cm:
0.20
50 cm:
0.50
1 m:
1.00
2 m:
2.00
4 m:
4.00
8 m:
8.00
16 m:
16.00
32 m:
32.00
:ACQuire:SMPinterval:VALue?
Function Queries the sampling interval.
Syntax
:ACQuire:SMPinterval:VALue?
Example
:ACQuire:SMPinterval:VALue? ->
:ACQ:SMP:VAL 8.0
:ACQuire:WAVelength
Function Sets the measured wavelength or queries the
current setting.
Syntax
:ACQuire:WAVelength {<NRf>}
:ACQuire:WAVelength?
<NRf> = 0.850um to 1.650um
(850E-9 to 1650E-9)
Example
:ACQuire:WAVelength 0.85um
:ACQuire:WAVelength 1650E-9
:ACQuire:WAVelength 1.650E-6
:ACQuire:WAVelength? ->
:ACQ:WAV 1550E-09
Description The wavelength that you can specify varies
depending on the model.