![YOKOGAWA AQ7277 Скачать руководство пользователя страница 22](http://html.mh-extra.com/html/yokogawa/aq7277/aq7277_user-manual_911224022.webp)
3-11
IM AQ7277-17EN
Before Programming
1
2
3
4
5
App
Index
3.5 Synchronization with the Controller
Achieving Synchronization
If you send the following program message when receiving averaging measurement
data, the data may be retrieved before the completion of the measurement.
ACQuire: AVERage: STARt; : WAVedata: DISPlay: SEND: ASCii?
In such case, the following method must be used to synchronize with the end of the
acquisition.
Using the STATus:CONDition? Query
The "
STATus:CONDition?
" command is used to query the contents of the condition
register.
Whether averaging measurement waveforms are being retrieved can be determined by
reading bit 1 of the condition register.
If bit 1 of the condition register is "
1
", waveforms are being retrieved. Otherwise, it is
stopped.(B
it 1 is "
0
".
)
Note
For details on the condition register, see section 4.4, "Condition Register."