4-6
IM AQ7277-17EN
4.3 Device-Specific Commands
:ACQuire:AVERage:COUNt?
Function Queries the current average count.
Syntax
:ACQuire:AVERage:COUNt?
Example
:ACQuire:AVERage:COUNt? ->
:ACQ:AVER:COUN 0
:ACQuire:AVERage:INDex
Function Sets the average count or queries the current
setting.
Syntax
:ACQuire:AVERage
:INDex {AUTO|E2_10|E2_11…….E2_20}
:ACQuire:AVERage:INDex?
Example
:ACQuire:AVERage:INDex AUTO
:ACQuire:AVERage:INDex E2_12
:ACQuire:AVERage:INDex? ->
:ACQ:AVER:IND E2_15
Description Setting the averaging count
Auto: AUTO
2E+10:
E2_10
2E+11:
E2_11
• • • • • •
2E+20:
E2_20
:ACQuire:AVERage:MODE
Function Sets the average mode or queries the current
setting.
Syntax
:ACQuire:AVERage
:MODE {HIREFLECTION|HISPEED}
:ACQuire:AVERage:MODE?
Example
:ACQuire:AVERage:MODE HIREFLECTION
:ACQuire:AVERage:MODE? ->
:ACQ:AVER:MODE HISPEED
Description High Reflection : HIREFLECTION
High Speed
: HISPEED
:ACQuire:AVERage:STARt
Function Executes the averaging measurement.
Syntax
:ACQuire:AVERage:STARt
Example
:ACQuire:AVERage:STARt
Description This command is valid while the averaging
measurement is only in progress.
ACQuire Group
The commands in this group deal with the waveform
acquisition. Change the measurement conditions through
remote control while the measurement is stopped. You
cannot change the measurement conditions while realtime
or averaging measurement is in progress. Only :REALtime:
STOP or AVERage:STOP is valid while the measurement is in
progress.
:ACQuire:ATTenuation
Function Sets the attenuation or queries the current setting.
Syntax
:ACQuire:ATTenuation {<NRf>|AUTO}
:ACQuire:ATTenuation?
<NRf> = 0.00 to 27.50 (steps of 2.5)
Example
:ACQuire:ATTenuation 2.5
:ACQuire:ATTenuation AUTO
:ACQuire:ATTenuation? ->
:ACQ:ATT 25.00
Description The selectable range varies depending on the
wavelength, distance range, and pulse width.
You cannot set the attenuation if the average
mode is high reflection.
You cannot set the pulse width if the distance
range is set to auto.
:ACQuire:AUTO:ATTenuation?
Function Queries the attenuation for AUTO.
Syntax
:ACQuire:AUTO:ATTenuation?
Example
:ACQuire:AUTO:ATTenuation? ->
:ACQ:AUT:ATT 0.00
Description The unit is set to m. (meter)
:ACQuire:AUTO:DRANge?
Function Queries the distance range for AUTO.
Syntax
:ACQuire:AUTO:DRANge?
Example
:ACQuire:AUTO:DRANge? ->
:ACQ:AUT:DRAN 20000
Description The unit is set to s. (second)
:ACQuire:AUTO:PWIDth?
Function Queries the pulse width for AUTO.
Syntax
:ACQuire:AUTO:PWIDth?
Example
:ACQuire:AUTO:PWIDth? ->
:ACQ:AUT:PWID 100E-09