5-26
IM AQ6373-01EN
Correcting the Resolution
1.
Press
SETUP
. The soft key menu for sweep condition settings appears.
2.
Press the
MORE 1/2
soft key.
3.
Press the
RESOLN CORRECT OFF ON
soft key. Each time you press the soft
key the setting toggles between ON and OFF.
Note
• Since the wavelength resolution function is set according to the monochromator slit width,
the setting resolution and actual resolution may not match. For this instrument, if the
resolution is set to 0.1 nm, the actual resolution will be 0.1 nm for a 450 nm wavelength, and
0.07 nm for a 850 nm wavelength.
• If the Resolution Correction function is turned ON, measured data is processed by the
software so that it matches the set resolution.
The Resolution Correction function is available when the resolution is set between 0.1nm
and 10.0 nm. The Resolution Correction function is only available when the horizontal axis
is in wavelength mode. It cannot be used in Frequency mode.
5.7 Wavelength (Frequency) Resolution Settings