3-16
IM AQ6373-01EN
Setting the Calibration Wavelength Value of External Light Source
1.
Press
SYSTEM
. The soft key menu is displayed.
2.
Press the
WL CALIBRATION
soft key.
3.
Press the
EXECUTE LASER
soft key. A screen for specifying the wavelength of
the external light source is displayed.
When entering a calibration wavelength value on the instrument
4.
Select the laser wavelength value using the rotary knob or the arrow keys. The
allowed wavelength range is 350–1200 nm.
5.
Press
nm/ENTER
. The wavelength value is set.
6.
Press the
EXECUTE
soft key. Wavelength calibration is executed. The previous
screen is displayed again after the calibration process ends.
7.
Press the
CANCEL
soft key during wavelength calibration to cancel the
wavelength calibration process.
3.6 Wavelength Calibration