10-12
IM AQ6373-01EN
No. Message
Cause of Warning
122 USB Storage is write protected
The USB storage device is write protected.
123 File not found
The specified file cannot be read because it has not been found. Or,
the file does not exist on the disk.
124 Illegal directory name
Directory cannot be created because the name is not valid.
125 Illegal file name
A file cannot be saved due to an incorrect file name.
126 Directory already exists
Directory cannot be created because a directory of the same name
already exists.
128 File is write protected
The file cannot be re-written or deleted because its property has
been set to READ ONLY.
129 Storage full
The file cannot be saved, because the hard disk or the USB storage
device is full.
130 Directory full
No files can be created because the directory area is full.
131 No data
There was no data to be saved, although an attempt was made to
save the file.
132 File is not a trace file
The file cannot be read because it is not a waveform file.
133 File is not a program file
The file cannot be read because it is not a program file.
134 File is not a data file
The file cannot be read because it is not a data file.
135 File is not a settings file
The file cannot be read because it is not a settings file.
137 File is not a template file
File cannot be loaded since it is not a template file or there is a
problem with the format of the template data.
138 Cannot copy
The file copy cannot be executed because the “copy from” file and
the “copy to” file have the same name.
140 No paste possible
The paste operation cannot be executed during the editing of
programs because the number of blank rows is not sufficient.
141 No merge possible
It is not possible to execute a merge during the editing of programs
because the merged result would exceed the maximum number of
rows.
142 WL calibration failed
Calibration cannot be executed because the level of a light source
is not sufficient during the wavelength calibration or because
wavelength deviations exceed the calibration range.
143 Optical Alignment failed
Alignment adjustment cannot be executed because the level of a
light source is not sufficient during the alignment adjustment.
144 Go/No go judgment stopped
Go/No go judgment stopped because template data reading or
AUTO sweep was carried out with Go/No go judgment function set
to ON.
145 No template data
Go/No go judgment or template data display was attempted to be
made with no template data provided.
160 Printer paper empty
There was no printer paper, although an attempt was made to use
the internal printer for printing.
161 Printer head up
It was not possible to execute printing because the head-up lever of
the internal printer was raised.
164 Printer Head Temperature error
The temperature of the built-in printer’s printer head is abnormal
when printing.
170 Illegal character
An illegal character was entered for a network name.
171 Illegal address
An illegal address was set to an IP address.
172 This version is not compatible
Attempted to load an incompatible update file.
No. 200 - 299: Warnings for hardware failures
200 Fan motor stopped!
The fan motor (main unit) stopped. Automatically shuts down after
10 seconds from occurrence.
201 Calibration data failed!
Started up in emulation mode because there was a problem with the
instrument’s calibration data.
202 Fan motor stopped!
The fan motor (CPU) stopped. Automatically shuts down after 10
seconds from occurrence.
205 Internal communication error!
An abnormality occurred during instrument-internal communication.
206 Internal communication error!
An abnormality occurred during instrument-internal communication.
207 Internal Temperature Over!
Internal temperature is abnormally high. Automatically shuts down
after 10 seconds from occurrence.
210 Internal Temperature warning!
Warning for rising internal temperature.
10.10 Warning Display Function