2-4
IM AQ6150B-02EN
Installation Conditions
Install the instrument in a place that meets the following conditions.
Well-Ventilated Location
There are inlet holes on the bottom side of the instrument. There are also exhaust holes on the rear
side. To prevent internal overheating, allow for enough space around the instrument (see the figure
below), and do not block the inlet and exhaust holes.
20 cm or more
When connecting cables, allow for enough space, above and beyond the space shown in the figure
above, to carry out the procedure.
Ambient Temperature and Humidity
Ambient temperature
5 to 35 °C
Ambient humidity
20 to 85% RH (no condensation)
Note
Condensation may form when moving the instrument from a low temperature or humidity environment to a
high temperature or humidity environment, or when there is a sudden change in temperature. In such cases,
before you use the instrument, allow it to adjust to the ambient temperature.
If you transport the instrument in its packing box, to prevent condensation, allow it to adjust to the new
ambient temperature before taking it out of the box.
Flat, Even Surface
Install the instrument on a stable surface that is level in all directions and that is not slippery. If you
use the instrument in an unstable location or on a slippery surface, the instrument may fall off its
platform or hit other devices due to the vibration that the instrument produces during operation.
If the instrument is installed in a horizontal position, the supplied rubber stoppers can be attached to
the feet at the rear of the instrument to prevent the instrument from sliding.
Rubber foot
A9088ZM
Foot at the rear of the
bottom of the instrument
2.2 Installing the Instrument