4-4
IM AQ6150B-02EN
5.
Press the
CONTINUE
soft key.
A list of update firmware files and an update execution setup menu appear. The message “Please remove
USB storage device” appears.
6.
Remove the USB storage device. If the
instrument
is connected to a network, disconnect the
network cable from the
instrument
.
The soft key for executing the update operation becomes available.
Note
Do not reconnect the network cable that you removed from the instrument ETHERNET port until the updating
is complete.
Species to load
from a USB
storage device
Display the
system
information.
Load the update
firmware.
Executes updating
A list of update files
Specifies to load
from an external
PC over the
network
7.
Press the
EXECUTE
soft key.
The instrument automatically restarts and starts updating. During updating, the screen shows that updating
is in progress. When updating is complete, the instrument automatically shuts down. This completes the
update procedure. Turn the power on, and the instrument will start normally.
4.2 Updating the Firmware