PON Power Meter (-PPM suffix code)
1
2
3
4
5
6
7
8
9
10
11
12
Index
App
Skipping
If you specify Skip for a core number that you don’t need to measure, its data will not be saved. By
setting which cores to skip beforehand, you can avoid accidentally saving unnecessary data.
Saving Data
You can save up to three sets of data in the save area of the specified core. When the signal direction
is from the OLT to the ONU, you can save the data for two wavelengths at the same time.
For details about saving the data to a file, see section 9.2.
Deleting Data
You can delete the data for individual core numbers in a save area or delete all the data at once.
Starting Core Number
You can set the starting core number to a value within the following range.
1 to 9900
Tape Number Type
You can set the tape number type to one of the following options.
Off, a-b(2), a-c(3), a-d(4), a-e(5), a-f(6), a-g(7), a-h(8)
Number of Fibers or Tapes
You can set the number of fibers or tapes to a value within one of the following ranges.
Tape Number Type
Number of Fibers or Tapes
Off
10 to 100 fibers
a-b(2)
10 to 50 tapes
a-c(3)
10 to 33 tapes
a-d(4)
10 to 25 tapes
a-e(5)
10 to 20 tapes
a-f(6)
10 to 16 tapes
a-g(7)
10 to 14 tapes
a-h(8)
10 to 12 tapes
4.3 Selecting and Saving Core and Tape Numbers