IM 1E10C1-E
5-8
5. FUNCTION AND DATA SETTINGS
Normally, these are all required settings. The following settings are made depending on the applications
that are used.
The pulse output calculation can be set in parameter No.
“n01” by selecting flow rate or flow rate value after
damping.
In case the pulse output transistor should be OFF
ACTIVE, please change the parameter No. in “n02” to
“01”.
After setting the number changing mode, the unit of pulse
weight is set in parameter No. “F01”.
When pulses are to be output per some liter with the flow
rate span of
■ ■
m
3
/h, select “02” since L = 10
-3
×
m
3
Set the pulse weight “10” in parameter No. “F02”.
Set the low cut range nearby 0% in parameter No.
“F03”.
Setting Range: 0 to 100% (of span)
The pulse width can be set in parameter No. “F04”.
Display
Description
Volume Unit
Code
00
01
02
03
04
05
06
Volume unit in that for the flow rate span
×
10
-9
Volume unit in that for the flow rate span
×
10
-6
Volume unit in that for the flow rate span
×
10
-3
Volume unit in that for the flow rate span
×
1
Volume unit in that for the flow rate span
×
10
3
Volume unit in that for the flow rate span
×
10
6
Number of pulses output per second at 100%
output
Pulse Width
Code
00
01
02
03
04
05
06
50%DUTY (Max. 1000p/s Min. 0.0001p/s)
0.5ms
(Max. 1000p/s Min. 0.0001p/s)
1ms
(Max. 500p/s Min. 0.0001p/s)
20ms
(Max. 25p/s Min. 0.0001p/s)
33ms
(Max. 15p/s Min. 0.0001p/s)
50ms
(Max. 10p/s Min. 0.0001p/s)
100ms
(Max. 5p/s Min. 0.0001p/s)
Default: 01
Default: 00
Default: 00
Default: (3%)
Default : 06
Default : 0
(Flow rate value after damping)