IM 1E10C1-E
11-10
11. PARAMETER LIST
Tl set value
0 to 999999
0
W
Sets forward direction totalization preset value.
E05
Total set
(00) Inhibit
Inhibit
W
Inhibits or enables forward direction totalization and reverse
E04
(01) Enable
direction totalization from being preset and reset, respectively.
Total switch
0 to 999999
0
W
Sets switching level when using status output function as
E06
totalization switch function.
Reverse total
0 to 999999
0
W
Displays and resets reverse direction totalizaed flow rate.
A31
Diff. total
–999999 to 999999
0
R
Displays differential totalized flow rate between forward and
A32
reverse.
(Diff. total = Totl – Reverse total)
Analog low cut
0 to 10%
0
W
Sets low cut width of current output.
G01
Analog low lmt
–20 to 100%
–20
W
Sets low limit of current output.
G02
Analog hi lmt
0 to 120%
120
W
Sets high limit of current output.
G03
PV AO Alrm typ
—
Hi
R
Displays burn out direction of current output.
—
Analog alarm out
(00) 2.4 mA or less
21.6 mA
W
Selects current output during alarm occurrentce.
b11
(01) 4.0 mA
or more
(02) Hold
(03) 21.6 mA or more
Poll addr
0 to 15
0
W
Sets polling address when multidrop mode.
—
Num req preams
5
5
R
Displays number of request preambles.
—
Burst mode
Off
Off
W
Selects the mode for the burst mode functionality.
—
On
Burst option
PV
PV
W
Selects sending items (instantaneous flow rate, output in %,
—
% range/current
totalization value and/or current output) when burst mode.
Process vars/crnt
Output function
(00) Pulse out
Pulse out
W
Selects status output functions.
b10
(01) Alarm out
(02) Bi direction
(03) Auto 2 ranges
(04) Low alarm
(05) Total switch
Reverse span
0.0001 to 30000
1.0000
W
Sets span for reverse direction.
b30
Bi direction hys
0 to 10%
2
W
Sets hysteresis when change in flow direction.
b31
Forward span2
0.0001 to 30000
1.0000
W
Sets the second span for forward direction.
b33
Auto range hys
0 to 15%
10
W
Sets hysteresis when change in Auto 2 ranges.
b34
Low alarm
–10 to 110%
–10
W
Sets low limit alarm.
b36
Low alarm hys
0 to 10%
5
W
Sets hysteresis for low limit alarm.
b37
Pulse unit
(00) n UNIT/P
PULSE/s
W
Selects pulse rate unit.
F01
(01) u UNIT/P
(“UNIT” in the selection on the left represents the “Base vol
(02) m UNIT/P
unit” selected.)
(03) UNIT/P
(04) k UNIT/P
(05) M UNIT/P
(06) PULSE/s
Name
Data Range, Units
Default
Value
R/W
Description
Indicator
of SE