IM 1E10C1-E
4-5
4. BASIC OPERATING PROCEDURES
4.2.2
Flow Rate Display Mode
• Flow Rate Display Mode indicates flow rates and totalization data. ADMAG SE can
indicate 12 types as follows.
DISPLAY ITEM
CONTENTS
BT200 SETTING
HART
Communicator
INDICATOR
Parameter
No.
Code
No.
Flow Rate (%)
Flow Rate
Actual Flow Rate
Actual Flow Rate
Totalization Values
Reverse Flow Rate
Totalization Values
Differential Between the
Forward and Reverse Flow
Totalization Values
Alternate Display of Flow Rate
(%) and Forward Flow Rate
Totalization Values
Alternate Display of Actual
Flow Rate and Forward Flow
Rate Totalization Values
Alternate Display of Actual
Flow Rate and Forward Flow
Rate(%)
Alternate Display of Flow Rate
(%) and Reverse Flow
Totalization Values
Alternate Display of Forward
Flow Rate and Reverse Flow
Totalization Values
Alternate Display of Flow Rate
(%) and Differential Flow
Totalization Values
Alternate Display of Forward
Flow Rate and Differential
Flow Totalization Values
D01: DISP SELECT
RATE(%)
D01: DISP SELECT
RATE
D01: DISP SELECT
FOR. TOTAL
D01: DISP SELECT
REV. TOTAL
D01: DISP SELECT
DIF. TOTAL
D01: DISP SELECT
RATE %/FOR TTL
D01: DISP SELECT
RATE/FOR TTL
D01: DISP SELECT
RATE/RATE %
D01: DISP SELECT
RATE %/REV. TTL
D01: DISP SELECT
RATE/REV. TTL
D01: DISP SELECT
RATE %/DIF. TTL
D01: DISP SELECT
RATE/DIF. TTL
Disp Select
PV % rnge
Disp Select
PV
Disp Select
Totl
Disp Select
Rev. totl
Disp Select
Diff. totl
Disp Select
PV % rnge/Totl
Disp Select
PV/Totl
Disp Select
PV % rnge/PV
Disp Select
PV % rnge/Rev. totl
Disp Select
PV/Rev. totl
Disp Select
PV % rnge/Diff. totl
Disp Select
PV/Diff. totl
Instantaneous flow rate is displayed
within a range of -8 (or -108%) to
108% for the span.
The actual meter rate of instataneous
flow rate is displayed. (See NOTE)
The decimal place is the same as for
the span setting. However, since a
decimal point set at the least
significant bit cannot be displayed.
Display actual flow rate totalization
value
Display reverse flow rate totalization
value
Differential totalization between
forward totalization and reverse
totalization is displayed.
Display alternately between display
of “RATE(%)” and “FOR. TTL”
every 4 second interval.
Display alternately between display
of “RATE” and “FOR. TTL”.
Display alternately between display
of “RATE” and “RATE(%)” every 4
second interval.
Display alternately between display
of “RATE(%)” and “REV. TTL”
every 4 second interval.
Display alternately between display
of “RATE” and “REV. TTL” every 4
second interval.
Actual flow rate (%) and differential
between forward and reverse
totalization values are indicated
alternately every 4 sec.
Actual flow rate and forward and
reverse totalization values are
indicated alternately every 4 sec.
d01
d01
d01
d01
d01
d01
d01
d01
d01
d01
d01
d01
00
01
02
03
04
05
06
07
08
09
10
11
• The display can set the units by setting parameter No.“b04” and “b05”.
• When special display of flow rate is specified in parameter “D10 flow user unit”, this
special display has higher priority in displaying than actual flow rate.
• Those 12 types function can be selected and set by changing into the Setting Mode.
(Please refer to 4.2.3 Setting Mode.)
• Those 12 types function can be selected by using infrared switches, Brain Terminal,
or HART communicator. (For using BT, please refer to Chapter 6. OPERATION VIA
BRAIN TERMINAL(BT 200). For using HART communicator, Chapter 7. OPERA-
TION VIA HART Communicator.)
NOTE