34
12.2
Menu
→
Configuration
→
Analysis
SUB MENU
FUNCTION
Fuel
The fuel menu allows the user to select the type of fuel to be used during
analysis.
Condensation
The burner efficiency figure when condensation takes place is influenced by
atmospheric pressure and humidity of the combustion air. As the atmospheric
pressure is an unknown, the operator is asked to enter a related parameter, i.e.
the altitude of the place above the sea level. The pressure is then calculated
from the atmospheric conditions. In the calculations the value of 101325 Pa is
assumed as atmospheric pressure at sea level. The air relative humidity input is
calculated at the combustion air temperature as measured from the CA502. If
the relative humidity value is unknown the operator is recommended to enter
50% for the value.
O
2
reference
This mode allows the user the ability to set the oxygen percentage level to which
pollutant emission values detected during analysis will be referenced.
Measure units
This submenu allows the user to modify the units of measurement for all the
analysis parameters.
Autozero
This sub menu is used to modify the auto zero cycle duration and the duration of
the sensor cleaning cycle.
Air temp.
This submenu
is used to acquire or manually enter the combustion air tempera-
ture.
15/06/18 10:00
ANALYSIS
►
Fuel
Condensation
O
2
reference
Measure units
Autozero
Air temp.
KEY
FUNCTION
Returns to the previous screen.
Selects the available parameters.
Enters in the selected parameter.
ESC
>
>
OK