6.2
Measurement and Accuracy Ranges
MEASUREMENT
SENSOR
RANGE
RESOLUTION
ACCURACY
RESPONSE
TIME T90
O
2
Electrochemical
sensor
0 .. 21.0% vol
0.1% vol
±0.2% vol
<20 sec.
CO
high H2 immunity
with NOx filter
Electrochemical
sensor
0 .. 4000 ppm
1 ppm
±20 ppm 0 .. 400 ppm
±5% measured value 401 .. 4000 ppm
<30 sec.
NO
Electrochemical
sensor
0 .. 2000 ppm
1 ppm
±5 ppm 0 .. 100 ppm
±5% measured value 101 .. 1000 ppm
±10% measured value 1001 .. 2000 ppm
<40 sec.
NOx
Calculated
CO
2
Calculated
0 .. 99.9% vol
0.1% vol
PI
*
(CO/CO
2
ratio)
Calculated
0.01%
Air temperature
TcK sensor
-
4 .. 248 °F (
-
20 to
120
0
C)
0.2 °F (0.1 °c )
±1 °F (±0.5 °C)
<30 sec.
Flue gas
temperature
TcK sensor
-
4 .. 1472 °F (
-
20 to
800
0
C)
0.2 °F (0.1 °c )
±1 °F (±0.5 °C) 32 .. 212 °F (0 to 100°
C)
±1% measured value 213 .. 1472 °F (101
to 800°C)
<30 sec.
Pressure
(draft & differential)
Piezoelectric
sensor
0.004 “
±1% measured value
-
40 .. +0.8
±0.008
-
0.8 .. +0.8
±1% measured value
+0.8 .. +80
<10 sec
Differential
temperature
Calculated
32 .. 2282 °F
(0 to 1250°C)
0.2 °F
Air index
Calculated
0.00 .. 9.50
0.01
Excess air
Calculated
0 .. 850 %
1 %
Stack loss
Calculated
0.0 .. 100.0 %
0.1 %
Efficiency
Calculated
0.0 .. 100.0 %
0.1 %
Efficiency
(condensing)
Calculated
0.0 .. 120.0 %
0.1 %
* The Poison Index ratio (P.I.) is a reliable indicator of a boiler or burner good operation. It only takes a simple flue gas test to determine
whether or not a service is needed to fix the system.