55
15.2
Sensor types and relevant positioning
S1
S2
O
2
Sensor
Part n. 68619
CO Sensor with NOx filter
0
-
4000ppm
Contact us for sensor
CODE
POSITION
15.3
Gas sensors life
The gas sensors used in this instrument are electrochemical: thus, when the relative gas is detected, a chemical
reaction takes place inside them that generates an electrical current.
The electrical current acquired by the instrument is then converted into the corresponding gas concentration.
Sensor life is strongly related to the consumption of the reagents within.
Sensor characteristics diminish as the reagents are consumed and when these have been used up completely
the sensor must be replaced. The sensors must be recalibrated on a regular basis to assure measuring accuracy:
recalibration can only be performed by a qualified YELLOW JACKET
®
service center.
Table 15.4 illustrates the characteristics inherent to each sensor.
GRAPHICAL DISPLAY OF ARRANGEMENT
15/06/22 10:00
SENSOR
►S1:O2 OK
S2:CO OK
S3:
——
-
——
-
SENSORS ARRANGEMENT INSIDE
THE SENSORS COMPARTMENT
15.0
SENSORS
15.1
Sensors arrangement
POSITION
S1 O
2
POSITION
S2 CO
CODE
MEASURED GAS
AVERAGE LIFE
RECALIBRATION
O
2
Sensor
Part n. 68619
O
2
Oxygen
24 months
not necessary
CO Sensor with NOx
H2 immunity
0
-
4000ppm
Contact us for sensor
CO
Carbon Monoxide
24 months
Yearly
15.4
Gas sensors life table