yellow jacket 40870 Скачать руководство пользователя страница 30

27 

 

Chapter 8:  

Troubleshooting Guide 

 

 

 

 

 

Model(s)

Problem

Possible Cause(s)

Possible Solution(s)

Check backlight auto-dimming timer (P51-870 only)

Tap power button or anywhere on display to power on backlight  (P51-870 only)

Device is not powering on

Make sure manifold is sufficiently charged

Screen is damaged

Contact technical support
Tap power button to power on backlight

Check backlight auto-dimming timer in device settings

Screen is damaged

Contact technical support

Display not calibrated properly

Recalibrate display (see User Manual/Quick Start Guide)

Display is damaged

Contact technical support

Device is connected to PC

Disconnect from PC

Buttons/overlay damaged

Contact technical support (P51-860 only)

Device is connected to PC

Disconnect from PC

Manifold is under pressure

Make sure manifold is vented to atmosphere

Pressure transducers damaged

Contact technical support

Temperature clamp barrel connector not 

fully seated in back of manifold

Check temperature clamp connections on back of manifold

Temperature clamp/cable damaged

Call technical support

Temperature clamp jacks damaged

Call technical support

Clamp not properly attached to system

Check clamp connections to system

Low-Side and High-Side readings reversed

Make sure T1 is attached to system low-side, T2 to system high-side

Wrong PT readings mode selected

Make sure Tsy mode is selected (P51-860 only)

Vacuum probe barrel connector not fully 

seated in back of manifold

Check vacuum probe connections on back of manifold

Vacuum probe/cable damaged

Call technical support

Vacuum probe jacks damaged

Call technical support

Probe not properly attached to system

Check vacuum probe connections to system

Vacuum probe plugged into wrong jack

Make sure vacuum probe is plugged into A1 or A2 jacks (P51-870 only)

Backlight has turned off

Screen has dimmed

Screen is auto-dimming, backlight is off

Pressure Transducers won't zero

P51-860 

P51-870

P51-860

P51-860 

P51-870

Screen does not display anything

P51-860 

P51-870

Device not responding to button press

Screen not responding to touch

P51-860 

P51-870

Table 8-1: P51 Troubleshooting

P51-870

Vacuum reading incorrect/not displaying

Temperature reading incorrect/not 

displaying

P51-860 

P51-870

Содержание 40870

Страница 1: ...ese Instructions do not cover the manifold attached to the instrument For instructions on use of the TITAN 4 Valve Manifold please visit https yellowjacket com product titan 4 valve test and charging...

Страница 2: ...des Menu 9 Pressure Temperature Mode 9 Overview 9 Interpreting the Pressure Analog Gauges 10 Interpreting the Pressure and Temperature Line Graphs 10 Operating Pressure Temperature Mode 11 Evacuation...

Страница 3: ...e Units 22 Temperature Units 22 Vacuum Units 22 Device Settings Menu 22 Brightness 22 Backlight Timer 22 Auto Off 23 Logging Rate 23 Ch 5 YJACK View App Integration 24 Overview 24 Operation of the YJA...

Страница 4: ...oid personal injury or death follow these guidelines Most governments and legal authorities re quire that HVAC technicians be trained and certified in the safe and proper operation of HVAC tools such...

Страница 5: ...damaged thermocouple or vacuum sensor could allow an electrical path across the instrument and over wet hoses Keep all interconnected equipment clean organized and in proper condition Do not use the i...

Страница 6: ...and release the power button located at bottom center of the display The YJ logo will appear briefly on startup The current firmware version and Serial Number are displayed at the bottom left and righ...

Страница 7: ...tions on touchscreen calibration see page 21 Connecting and Using the Temperature Clamps Two 67010 temperature clamps pictured in Figure 6 are supplied with the P51 870 TITAN and can be used to monito...

Страница 8: ...distance from the vacuum pump such that it does not disturb the vacuum measurements Interpreting the Battery Life Indicator This instrument utilizes a 2000mAhr rechargeable lithium ion battery It is e...

Страница 9: ...s equipped with an adjustable backlight and backlight auto dimming capabilities When the backlight dims during use a touch anywhere on the screen or a quick press of the power button can be used to wa...

Страница 10: ...nvention preventing duplicates See Figure 15 for example of stored datalog files Ex 50th week of 2017 with device serial number ending in 127 4th datalog file 750127AD These files cannot be modified w...

Страница 11: ...to a larger time increment the LED will not flash blue as often Flashing Red When the device reaches low charge or falls below 10 battery charge the LED will flash red every time a datalog sample is...

Страница 12: ...icon in the upper right hand corner Pressure Temperature Mode Overview The P51 870 TITAN Pressure Temperature mode can be used to accurately diagnose and service any system with compatible refrigerant...

Страница 13: ...s displayed in a digital format at the center of each blue and red analog gauge respectively Each gauge needle adjusts in real time and can be used to observe fluctuations in pressure as if utilizing...

Страница 14: ...e at any time by tapping anywhere within the graph Operating Pressure Temperature Mode While on any modes screen or the modes menu tap the settings icon in the upper right hand corner to access the ge...

Страница 15: ...l Settings Button System vacuum pressure is displayed in real time as a digital representation as denoted by Pvac analog as represented by the green analog gauge and a graphical format as a logarithmi...

Страница 16: ...ding Use the Target and Hold buttons to set the target vacuum level and vacuum hold timer respectively Pressing each of these buttons will cycle between a range of available values with the last value...

Страница 17: ...igerant and non condensables can cause the vacuum level to rise in a system giving a false positive during a leak test when no leak is present Pressure Hold Mode Pressure Decay Test Overview The P51 8...

Страница 18: ...ear scale with minor divisions displayed as small dots and major divisions displayed as larger dots respectively The current system pressure is displayed in a digital format at the center of the analo...

Страница 19: ...s measured as a percent of the initial system pressure the system can drop before the P51 870 TITAN deems the pressure decay test a failure To adjust the allowable change percent drag the slider bar u...

Страница 20: ...will display a FAIL notice Figure 24 and the buzzer will sound indicating the test is complete To silence the buzzer exit the pressure decay test by selecting the modes button If the pressure decay t...

Страница 21: ...gital analog vacuum gauge as evacuation mode The pressure rise gauge operates on a logarithmic scale from atmosphere ATM to 0 microns A digital reading and currently selected vacuum units are displaye...

Страница 22: ...desired settings have been set select the continue button to advance to the pressure rise test The back button can be used to return to the previous evacuation screen After selecting the continue but...

Страница 23: ...gs menus tap on the appropriate button The exit button in the bottom left corner can be used to return to the previous screen at any time Zeroing the Pressure Transducers To ensure accurate pressure r...

Страница 24: ...rigerant Settings Menu The refrigerant settings menu shown in Figure 31 can be accessed either by tapping on the Change Refrigerant button in the general settings menu or by tapping the button in the...

Страница 25: ...the Device Settings button in the general settings menu To adjust the display brightness tap and drag the slider bar next to brightness until the desired brightness is displayed To adjust the backligh...

Страница 26: ...t the logging rate to the lowest possible time interval 1s but this may result in extremely large datalog files making it difficult to analyze the information Be sure to set the logging rate to an app...

Страница 27: ...ce Navigate to the YJACK VIEW App and make sure the manifold is powered on The P51 870 TITAN should appear in the available devices menu as the name of the device followed by the serial number Ex YJP5...

Страница 28: ...dirty generously soak a soft cloth with warm soapy dish soap water and place the cloth for a couple of minutes over the display to loosen any stubborn dirt Wipe off excess water with a clean less damp...

Страница 29: ...acklight Size Approx 7 5 x 8 5 x 4 25 Weight 2 63 lbs Table 7 2 Instrument Specifications Working Pressure 0 700 psia 48 3 bar Pressure Sensing Resolution 0 1 psi 0 1 bar 1 kPa 0 001 MPa 0 01 kg cm 2...

Страница 30: ...d Temperature clamp cable damaged Call technical support Temperature clamp jacks damaged Call technical support Clamp not properly attached to system Check clamp connections to system Low Side and Hig...

Страница 31: ...28...

Отзывы: