1. SAFETY PRECAUTIONS
Warning and Cautions
Warning
DO NOT adjust pressure reduction valve.
The pressure reduction valve has a default water sampling rate of approximately 0.8L/min
(base water pressure: 0.30MPa, pressure reduction valve: 0.20MPa, throttle valve:
0.60MPa, water temperature: 20°C without membrane filter), as set at factory before
shipping. Do not attempt to adjust it.
Altering any of the factory settings may cause unit to leak or have decreased flow rate.
DO NOT turn ELB OFF.
Water is routinely flushed through the system, even while water samples are not being
drawn. Be advised that if the ELB is turned
“OFF” for an extended period of time, water
retained in the system cannot be flushed through, resulting in accelerated deterioration of
the reverse osmosis (RO) membrane degraded water purity.
DO NOT shut off water supply.
Water is routinely flushed through the system, to protect the RO membrane and maintain
water quality. The water supply tap should normally be left open to keep the flush function
enabled. Be advised that the flush function produces normal water discharge from system.
Prepare accordingly.
Caution
DO NOT operate equipment during thunderstorms.
In the event of a thunderstorm, terminate operation and turn off main power switch (ELB)
immediately. A direct lightning strike may cause damage to equipment, or result in fire or
electric shock.
Power loss.
Although unit naturally stops when a power failure occurs, turn main power switch (ELB)
“OFF” for safety. Follow procedures under “Turning power ON” (P.24) after power is
restored.