44
8. TROUBLESHOOTING
Malfunctions
More on leak detection
error
●
Confirmation
1. Confirm that the water supply connection coupler and quick couplers on ion exchange resin
cartridge, pre-process cartridge and the Reverse Osmosis (RO) membrane are securely
connected.
* If leaks are discovered coming from unit plumbing system, turn ELB off immediately, shut off
main water supply tap and contact local dealer or sales office for assistance.
2. Confirm that water has not dripped on leak sensor while replacing a cartridge or other wear item.
●
Solution procedures
(1) Turn the ELB
“OFF(
O
)
”.
(2) Isolate leak. Leak sensors are located at the two positions shown below:
・
Front side - visible when the door is opened.
・
Rear of main unit
(3) Remove leak sensor from main unit, thoroughly wipe all moisture from electrode and from
interior of unit. Allow them to dry completely.
(4) When dry, be sure to replace leak sensor(s) in original position.
(5) Replace door and close.
(6) Turn ELB
“ON (
|
)
” and press the POWER key to resume normal operation.
*
ALWAYS shut main water supply tap off BEFORE attempting to disassemble system
plumbing (i.e. to repair a leak).
Leak sensor