6. MAINTENANCE PROCEDURES
Wear Item Replacement Alert Reset
The useful life of each replaceable cartridge is determined by cumulative water volume it can reliably
filter within a predicted timeframe.*1 The WE200 unit counts the cumulative water volume filter time
(in hours) for each cartridge and generates notifications to the user when replacement is needed
(See display descriptions on P.41). When replacing filter cartridges, reset cumulative water volume
hours using the procedures outlined below.
Resetting replacement notifications/alerts
①
Turn the ELB
“ON (
|
)
”.
②
Press POWER.
・
For single item alert
③
Pressing and holding CLR sounds an
acceptance tone and resets the notification
so that normal operation may be resumed.
・
For multiple item alert
③
Pressing and holding CLR sounds acceptance
tone and resets the
notification.
④
Pressing and holding CLR again sounds
acceptance tone and resets the
notification.
⑤
Repeat to clear any other alert items so that
normal operation may be resumed.
*
Reset the alerts ONLY after replacing items for which alerts were generated.
*
When multiple replacement alerts are generated, reset order is as follows. Replace all items
and clear all alerts before attempting to resume operation.
Reset order:
→
→
→
*
If premature replacement notifications/alerts are suspected, call for service.
*1 Useful life of ion exchange resin cartridge is determined by water quality.
②
③