23
4. Operating procedures
Operating method
of GB210
(
6) When the sample reaches the granule diameter you want,
sufficiently dry the samples, and turn the heater OFF.
When outlet temperature has decreased to 45
℃
or lower,
turn the blower OFF and choke spray pressure to 0.
※
To avoid a malfunction of the unit, do not turn the
blower OFF while the outlet temperature is 45
℃
or
more.
(7) Using the pulse jet switch, remove foreign objects off the
filter.
(8) Lower the stage with the lift DOWN switch and then turn the
POWER switch OFF.
Take out the flow layer chamber and collect the granulated
samples.
(9) Wash the containers according to the maintenance method
(P33 and following pages “Cleaning After Using”).