2. Before using this unit
Service receptacle capacity
Service receptacle capacity
Apply the 100V 2A or less service receptacle for this unit.
Connecting the service receptacle with its capacity over 2A blowouts the fuse, and the power
source to the service receptacle is shut down. For resetting this damage, replace the fuse in
the fuse holder on the right side of the back of the unit.
Applicable models
Mag mixer: MA series, M-21, MD series, MC800, MF800
Use a separate power supply for a unit with a heater and its total current exceeds 2A
Temperature output terminal
Temperature output signals at the outlet, the inlet during spray drying
(
GB210A
)
, and the inlet during
granulation
(
GB210B
)
are of 4-20mA for the measured temperature range of 40-240
℃
.
[
Current output of 4-20mA
:
Measured temperature of 40-240
℃
]
Conversion formula: Current output I (mA)=0.08 x (measured temperature T(
℃
)
+
10)
Measured
temperature
T(
℃
)
=
12.5×current output I (mA)-10
When you connecting to the voltage input of the recorder, connect a fixed resistor (shunt resistor)
of 600
Ω
or lower.
Inlet temp
(4~20mA:40~240℃)
Inlet temp(GRANULE)
(4~20mA:40~240℃)
Outlet temp
(4~20mA:40~240℃)
20mA
4mA
40℃
240℃
Measured
temp.
Current output
Содержание PULVIS MINI BED GB210B
Страница 39: ...37 8 When a trouble occurs Confirmation of GB210A manual and language selection display ...
Страница 40: ...38 8 When a trouble occurs Confirmation of GB210A manual and language selection display ...
Страница 41: ...39 8 When a trouble occurs Confirmation of GB210B manual and language selection display ...
Страница 42: ...40 8 When a trouble occurs Confirmation of GB210B manual and language selection display ...