19
4. Operating procedures
Operating method
of GB210B
Operating method of GB210B
(
GB210 main unit + GF200 mini bed
)
Granulation method
Refer to sample settings for an example case for the standard sample.
Sintered alumina
(
W.A.
♯
180
)
NET
300g
PVA
(
Polyvinyl alcohol
)
♯
500 NET
50g
Solid component density
5wt
%
(
1
)
The temperature controller at the upper left part of the
operation panel is used as the display and setting
device for inlet temperature and the lower temperature
controller, as the display for outlet temperature.
SWITCHING is not used when the mini bed is used.
(
2
)
Preliminary heating and setting of samples are
performed here.
(
2
)
-1) Install mini bed attachments following the above
procedures. Adjust the spray nozzle so that its tip will
protrude below by about 30mm from the end of the
pipe.
Set spray pressure to 0.01MPa.
(
2
)
-2) Turn the blower switch in the key panel ON.
(
2
)
-3) Adjust the left volume so that an appropriate air
amount is obtained while monitoring it on the air
amount monitor.
Example Air amount: approx.0.4m
3
/min
Содержание PULVIS MINI BED GB210B