26
6. HANDLING PRECAUTIONS
Warnings and Cautions
Circulation pump.
・
Never operate circulation pump dry. Damage or malfunction may result.
Operating unit without circulating water in the reservoir and circulation pump may cause
seizure of the pump and/or other problems.
・
Keep water reservoir covered and do not allow any debris to enter and remain in the
reservoir. Damage to the pump may result.
・
Do not operate with the circulating path completely closed.
・
Make sure that the flow rate of circulating water keeps at least 1 L/min.
Refrigerator abnormal stop
If room temperature and liquid temperature are high, or the intake dust filter is clogged,
increasing the load of the refrigerator. The refrigerator high-pressure cut-off switch may
operate and stop the refrigeration.
Calibration offset
If there is a deviation between actual temperature of the circulating water and temperature
reading, and adjustment is necessary, refer to "Calibration offset" (P.21) to perform
temperature correction.
Inspect regularly.
ELCB, in particular, is a key device in maintaining this unit safety, and must be
inspected/maintained regularly. See "Inspection and Maintenance" (P.27)