![Yamatake 900 Series Скачать руководство пользователя страница 226](http://html.mh-extra.com/html/yamatake/900-series/900-series_user-manual_905236226.webp)
Maintenance and Troubleshooting
Yamatake Corporation
7-26
ST3000 Smart Transmitter Series 900 Electronic Differential Pressure/Pressure Transmitter
7-5-2 :Erasing # Mark
Reset the calibrated data to the initial values (factory set data):
How to calibrate the low limit value, based on actual pressure
Step
Description
SFC screen
1
Set the actual pressure applied to
the transmitter to 0kPa.
2
Press the
key.
3
Press the
key.
4
Press the
keys.
• The initial value for the low
limit is displayed.
Step
Description
SFC screen
5
Press the
key.
6
Press the
key.
7
Press the
key.
Ensure that no pressure is applied to H.L.
DSTJ FIT-1234
READY...
LRV E
0%
LRV FIT-1234
0.0000 kPa #
G
SET
LRV FIT-1234
SET LRV?
ENTER
( Yes )
NON-VOL
LRV FIT-1234
WORKING...
LRV FIT-1234
0.0123 kPa #
Z
0
ACT PR
LRV FIT-1234
0 kPa #
ENTER
( Yes )
NON-VOL
LRV FIT-1234
0.0000 kPa #
COR- K
RECT
RESET
LRV FIT-1234
CORRECT LRV?