Instrument does not react to keystroke
10.3.4
Displaying the software version (meter)
Cause
Remedy
Operating condition undefined or
EMC load unallowed
• Processor reset:
Press the
<ENTER>
and
<On/Off>
key simultaneously
Cause
Remedy
E. g., a question by the service
department
• Switch on the meter.
During the display test, display
the software version with
<ENTER>
.