Operation
TruLab pH 1110
14
ba76155e01
03/2014
4.1.2
Display
4.1.3
Status information
1 Status information
2 Measured value (with unit)
3 Measured parameter
4 Sensor symbol (calibration evaluation, calibration interval)
5 Measured temperature (with unit)
8
8 8
8
8
1
8
8
8
°C
°F
UpH
mV/pH
LoBat
AutoCal DIN
AutoCal TEC ConCal
AR
TP
CalError
%
Time
3
2
4
5
1
Display
Meaning
[AutoCal TEC]
[AutoCal DIN]
Calibration with automatic buffer recognition, e.g. with the
buffer set: Technical buffers
[ConCal]
Calibration with any buffers
[CalError]
An error occurred during calibration
[AR]
Stability control (AutoRead) is active
[TP]
Temperature measurement active
[Time]
Setting of calibration interval
[LoBat]
Batteries are almost empty