TruLab pH 1110
What to do if...
ba76155e01
03/2014
39
10.1.3
Error message OFL, UFL
The measured value is outside the measuring range.
The measured value is obviously incorrect.
10.2
ORP
10.2.1
No stable measured value
Cause
Remedy
• The used buffer solutions do not
agree with the set buffer set
• Set different buffer set
or
• Use different buffer solutions
• Buffer solutions too old
• Use only once.
Note the shelf life
• Buffer solutions depleted
• Change solutions
Cause
Remedy
• Cable broken
• Replace the electrode
• Gel electrolyte dried out
• Replace the electrode
• Air bubble in front of the junction
• Remove the air bubble
• Air in the junction
• Extract air or moisten the junction
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause
Remedy
• Junction contaminated
• Clean the junction
• Pt-ORP electrode contaminated
• Clean the Pt-ORP electrode