Reset
TruLab pH 1110
34
ba76155e01
03/2014
8
Reset
You can erase the calibration values and reset (initialize) the measurement and
system settings.
8.1
Resetting the calibration values
1.
Press
<On/Off__>
to open the menu for the reset of the calibration data.
InIt.C
is displayed.
2.
Use
<
><
>
to display
no
or
YES
.
•
YES
: Reset the calibration values
•
no
: Retain the calibration values.
3.
Confirm
with
<ENTER>.
The menu is finished. The meter switches to the measuring mode.
Calibration values to be reset
8.2
Resetting the measurement and system settings
1.
Switch on the meter with
<On/Off>
.
The display test appears briefly on the display.
2.
During the display test, press
<M>
to open the menu for the reset of the
meter settings.
Init
is displayed.
3.
Use
<
><
>
to display
no
or
YES
.
•
YES
: Reset the meter settings.
•
no
: Retain the meter settings.
4.
Confirm
with
<ENTER>.
The settings are reset. The menu is finished.
The meter switches to the measuring mode.
The calibration values are reset to default. All other meter
settings are retained.
Recalibrate after performing a reset.
Calibration value
Default settings
Zero point (
ASY
)
0 mV (pH 7.000)
Slope (
SLO
)
-59.16 mV/pH (100 %)
Содержание YSI TruLab pH 1110
Страница 1: ...OPERATIONS MANUAL ba76155e01 03 2014 TruLab pH 1110 pH METER pH 1110...
Страница 50: ...Index TruLab pH 1110 50 ba76155e01 03 2014...
Страница 52: ...Contact Information TruLab pH 1110 52 ba76155e01 03 2014...
Страница 53: ......