pH value
TruLab pH 1110
18
ba76155e01
03/2014
5
pH value
5.1
Measuring
5.1.1
Measuring the pH value
1.
Connect the pH electrode to the meter.
2.
When measuring without temperature sensor: Temper the test sample or
measure the current temperature.
3.
If necessary, select the measured parameter (pH) with
<M>
.
4.
Immerse the pH electrode in the test sample.
The measured value is checked for stability (stability control).
The
[AR]
status display flashes.
5.
When measuring without temperature sensor: Enter the temperature of
the buffer with
<
><
>
.
6.
Wait for a stable measured value.
The
[AR]
display indicator no longer flashes.
Stability control (AutoRead )
During the measuring procedure, the stability control function is automatically
activated.
The stability control function (
AutoRead
) continually checks the stability of the
measured values in the monitored time interval. The stability has a consider-
able impact on the reproducibility of measured values. The
[AR]
display indi-
cator flashes until a stable value is measured.
To ensure the high measurement accuracy of the measuring sys-
tem, always measure with A current calibration (see section 5.2
C
8
4
2
2
9
9
6
pH
°C
TP
AR