Xylem YSI IDS ProOBOD Скачать руководство пользователя страница 8

8

9

6.  Locate the o-ring lubricant included with the new sensor cap.  

Apply a thin coat of o-ring lubricant to the installed o-ring. After 
application, there should be a thin coat of o-ring lubricant on the 
o-ring only. Remove any excess o-ring lubricant from the o-ring 
and/or probe with a lens tissue. 

7.  Clean the clear surface of the optical DO sensor window (figure 1) 

with a lens cleaning tissue.

8.  Remove the new sensor cap from its hydrated container and dry 

the inside cavity of the sensor cap with lens tissue. Make sure that 
the cavity is completely dry and clean before proceeding with the 
installation. Using a clockwise motion, thread the new sensor cap 
onto the probe assembly until it is finger-tight. The o-ring should 
be compressed between the sensor cap and probe. Do not over-
tighten the sensor cap and do not use any tools for the installation 
process.

9.  Clean the stir paddle and then reinstall it on the probe by pushing 

it straight into place.    

10.  Store the sensor in a BOD bottle with a small amount of water 

(approximately 40 ml).

11.  Enter the new Sensor Cap’s coefficients into the instrument. See the 

instrument’s manual for more information on entering coefficients. 

12.  After entering the new Sensor Cap coefficients, perform a 1-point 

DO calibration. 

REhYDRATING ThE SENSOR CAP  

The Sensor Cap must remain in a moist environment for proper 
operation; see Sensor Storage

 

for storage recommendations. If you 

inadvertently leave your sensor exposed to ambient air for a period 
of more than approximately 8 hours it may dry out. If the sensor cap 
is allowed to dry out, it is likely to drift slightly at the beginning of 
your next study unless it is rehydrated. If the cap dries out, you can 
rehydrate it by soaking the probe tip with the sensor cap installed in 
warm (room temperature) tap water for 24 hours. After rehydration is 
complete, perform a 1-point DO calibration and be sure to store the 
probe in a moist environment.   

SenSOr maintenance - temperatUre

You must keep the temperature portion of the sensor free of build 
up. Other than that, the sensor requires no maintenance. A soft bristle 
brush can be used to scrub the temperature sensor if needed. While 
cleaning, be sure to only scrub the temperature sensor and not the 
sensor cap. The sensor cap will be damaged if cleaned with anything 
abrasive.    

SenSOr StOrage

ShORT-TERM STORAGE  

When the OBOD is not in use, it must be stored in a moist environment, 
i.e., the sensor either immersed in water or in water-saturated air. If the 
sensor cap is allowed to dry out by exposure to dry air, it is likely to drift 
slightly at the beginning of its next use unless it is rehydrated. If this 
occurs, follow the rehydrating instructions in this manual.  

For short-term storage (<30 days), place the probe in a BOD bottle that 
contains a small amount of clean water (approximately 40 ml). This will 
provide a 100% water-saturated air environment.  

LONG-TERM STORAGE

For long-term storage (>30 days), remove the batteries from the 
instrument. Moisten the sponge in the protective plastic cap that was 
originally provided with the probe with clean water. Place the clear 
cap over the sensor with the sensor cap installed. Inspect the sponge 
every 30 days to make sure it is still moist. If you no longer have the  
protective cap, place the probe in a BOD bottle that contains a small 
amount of clean water (approximately 40 ml). Or, you can place the 
sensor with sensor cap installed directly in water in a BOD bottle.

Recommended Long-term Storage ambient 

temperature: -5 to 50°C (23 to 122°F)

T R O U B L E S h O O T I N G

DiSSOlveD Oxygen reaDingS

Erroneous dissolved oxygen readings typically indicate a need to 
clean the sensor cap, replace the sensor cap and/or recalibrate the 
instrument. 

First, verify that the instrument is measuring temperature accurately. 
Both the DO % and mg/L measurements are temperature dependent 
so if the temperature reading is incorrectly, the DO measurements will 
also be erroneous. 

If the DO% measurement is accurate but the mg/L value is incorrect, 
check the salinity compensation setting in the instrument. The salinity 
compensation value should be set to the salinity of the sample you are 
testing. Municipal wastewater is typically 0 ppt. 

Содержание YSI IDS ProOBOD

Страница 1: ...USER MANUAL Document 626471 ProOBOD IDS ProOBOD ...

Страница 2: ...Dissolved Oxygen 4 Temperature 5 Taking Measurements 5 Maintenance and Storage 5 Updating Instrument and Probe Firmware 6 Sensor Maintenance Dissolved Oxygen 6 Sensor Maintenance Temperature 8 Sensor Storage 9 Troubleshooting 9 Dissolved Oxygen Readings 9 Error Status messages 11 Probe Specifications 11 Accessories Parts List 12 Declaration of Conformity 13 Recycling 14 Contact Information 14 Orde...

Страница 3: ...ervice Center selected by YSI Repair or replacement will be made and the product returned transportation prepaid Repaired or replaced products are warranted for the balance of the original warranty period or at least 90 days from date of repair or replacement LIMITATION OF WARRANTY This Warranty does not apply to any YSI product damage or failure caused by 1 Failure to install operate or use the p...

Страница 4: ...nt It fits in standard USA BOD bottles standard Japanese BOD bottles and many other international BOD bottles A funnel adapter for fitting in standard UK BOD bottles and a switching power supply with 3 international outlet adapters are also included IDS ProOBOD YSI item number 626500 This probe works with a MultiLab IDS instrument It fits in standard USA BOD bottles standard Japanese BOD bottles a...

Страница 5: ... contact YSI Customer Service at 800 897 4151 1 937 767 7241 or environmental ysi com or contact the authorized YSI distributor where the instrument was purchased Remove the OBOD probe cable assembly from the shipping container and locate the instruction sheet that is included This instruction sheet is important because it contains the calibration coefficients for your sensor cap A new probe cable...

Страница 6: ...ift than traditional electrochemical sensors This increased stability means that the instrument may hold its calibration for many months However for the highest data accuracy YSI recommends verifying the instrument s calibration on a daily basis To verify the instrument s calibration place the sensor in a water saturated air environment BOD bottle with 40 ml of water and check to see that the DO m...

Страница 7: ...section of this manual for instructions on how to determine if the sensor cap needs replaced The replacement OBOD sensor cap is shipped in a humidified container and the package should not be opened until the cap is needed for use Once the sensor cap has been installed on the OBOD probe as described in this section it is important to keep the sensor in a 100 humid environment Therefore the OBOD se...

Страница 8: ...ature portion of the sensor free of build up Other than that the sensor requires no maintenance A soft bristle brush can be used to scrub the temperature sensor if needed While cleaning be sure to only scrub the temperature sensor and not the sensor cap The sensor cap will be damaged if cleaned with anything abrasive Sensor Storage Short term Storage When the OBOD is not in use it must be stored i...

Страница 9: ... carefully remove the o ring and contact YSI Technical Support to obtain a new o ring and or gasket Avoid using tools to remove the o ring as damage to the sealing surfaces could result Before reinstalling the sensor cap make sure that the cavity is completely clean and dry before proceeding with the installation If water is found dry the cavity with lens tissue Finally clean and dry the clear opt...

Страница 10: ...ms to Australian and New Zealand EMC requirements for C Tick and RCM marks It conforms to unintentional radiator requirements under FCC Part 15 and ICES 003 Class B Manufacturer YSI Incorporated 1725 Brannum Lane Yellow Springs OH 45387 USA Product Name Professional ODO Water Quality Instrument and Professional OBOD probe Model Numbers Instrument Accessory ProODO 626281 ProComm Saddle manufactured...

Страница 11: ...8 technical support Email environmental ysi com Mail YSI Incorporated 1725 Brannum Lane Yellow Springs OH 45387 USA Internet ysi com When placing an order please have the following available 1 YSI account number if available 2 Name and phone number 3 Purchase Order or Credit Card 4 Model Number or brief description 5 Billing and shipping addresses 6 Quantity Service Information YSI has authorized ...

Страница 12: ...106 87 25 73 653 6 87 2 871 4115 1254 86 25 43 646 86 1 861 4430 1350 85 25 13 638 4 85 1 851 4747 1447 84 24 84 630 8 84 1 841 5067 1544 83 24 54 623 2 83 1 831 5391 1643 82 24 24 615 6 82 1 821 5717 1743 81 23 94 608 81 06 811 6047 1843 80 23 64 600 4 80 05 800 6381 1945 79 23 34 592 8 79 03 790 6717 2047 78 23 04 585 2 78 02 780 7058 2151 77 22 74 577 6 77 01 770 7401 2256 76 22 44 570 0 75 99 ...

Страница 13: ...Item 626471 Rev C Drawing 626471 April 2013 2013 YSI Incorporated ...

Отзывы: