Xylem YSI IDS ProOBOD Скачать руководство пользователя страница 5

2

3

oxygen levels, and the elimination of frequent membrane/electrolyte 
changes and electrode servicing.  

For more information on the ProODO and MultiLab IDS instruments, 
including product specification, please visit ysi.com or contact YSI 
Technical Support at 800-897-4151 (+1 937-767-7241), environmental@
ysi.com.  

Reading the entire manual before use is recommended for an overall 
understanding of the probe’s features.  

Important:  The OBOD probes are designed for laboratory use, 

not field use. They are not waterproof and should never be 

immersed in a sample past the taper on the stem of the probe.

G E T T I N G   S T A R T E D

Throughout the manual, the term “probe” refers to the end of the cable 
where the sensor, stir motor and stirrer are located.  The term “sensor” 
refers to the Optical Dissolved Oxygen sensing portion of the probe 
assembly and the term “sensor cap” refers to the removable sensing 
cap that is replaced about once every 12 to 18 months (Figure 1).

Figure 1

Stirrer toggle switch, turns 

stir paddle on and off

 

Probe

 

 

Temperature sensor

 

Optical DO sensor

OBOD Sensor Cap, 

shown here removed 

from probe

 

 

Stir paddle

Unpacking the OBOD prOBe

Each OBOD probe and replacement OBOD sensor cap 

includes an instruction sheet with important information 

unique and specific to each individual cap.  

Carefully unpack the instrument and accessories and inspect for 
damage. Compare received parts with items on the packing list. If 
any parts or materials are damaged or missing, contact YSI Customer 
Service at 800-897-4151 (+1 937 767-7241) or [email protected] 
or contact the authorized YSI distributor where the instrument was 
purchased.

Remove the OBOD probe/cable assembly from the shipping container 
and locate the instruction sheet that is included.  This instruction sheet 
is important because it contains the calibration coefficients for your 
sensor cap. A new probe/cable assembly already has a sensor cap 
installed and the sensor cap coefficients are preloaded into the probe 
at the factory. The coefficients are automatically sent to the instrument 
when the probe is connected to the instrument. however, the probe 
instruction sheet should be saved in case you need to reference the 
coefficients in the future.

Preparing the probe for the first time:

1.  A new OBOD probe will have a protective cover with a moist 

sponge placed over the sensor cap to ensure the cap remains 
hydrated during shipment. Carefully remove this cover from the 
end of the probe by pulling it straight off the sensor. Save the cover 
along with the sponge to use later for long term storage of the 
probe. Long term storage is generally considered to be 4 weeks 
or longer. 

2.  Place the probe in a standard BOD bottle that contains a small 

amount of clean water (approximately 40 ml).  The sensor should 
not be immersed in water but rather in the air above the water.  
The purpose is to create a 100% water-saturated air environment 
for the sensor which is ideal for calibration and storage when not 
in use.    

It is important to always keep your sensor in a moist 

environment so the sensor cap does not dry out.  (See 

Maintenance and Storage for more information.)  

i

Содержание YSI IDS ProOBOD

Страница 1: ...USER MANUAL Document 626471 ProOBOD IDS ProOBOD ...

Страница 2: ...Dissolved Oxygen 4 Temperature 5 Taking Measurements 5 Maintenance and Storage 5 Updating Instrument and Probe Firmware 6 Sensor Maintenance Dissolved Oxygen 6 Sensor Maintenance Temperature 8 Sensor Storage 9 Troubleshooting 9 Dissolved Oxygen Readings 9 Error Status messages 11 Probe Specifications 11 Accessories Parts List 12 Declaration of Conformity 13 Recycling 14 Contact Information 14 Orde...

Страница 3: ...ervice Center selected by YSI Repair or replacement will be made and the product returned transportation prepaid Repaired or replaced products are warranted for the balance of the original warranty period or at least 90 days from date of repair or replacement LIMITATION OF WARRANTY This Warranty does not apply to any YSI product damage or failure caused by 1 Failure to install operate or use the p...

Страница 4: ...nt It fits in standard USA BOD bottles standard Japanese BOD bottles and many other international BOD bottles A funnel adapter for fitting in standard UK BOD bottles and a switching power supply with 3 international outlet adapters are also included IDS ProOBOD YSI item number 626500 This probe works with a MultiLab IDS instrument It fits in standard USA BOD bottles standard Japanese BOD bottles a...

Страница 5: ... contact YSI Customer Service at 800 897 4151 1 937 767 7241 or environmental ysi com or contact the authorized YSI distributor where the instrument was purchased Remove the OBOD probe cable assembly from the shipping container and locate the instruction sheet that is included This instruction sheet is important because it contains the calibration coefficients for your sensor cap A new probe cable...

Страница 6: ...ift than traditional electrochemical sensors This increased stability means that the instrument may hold its calibration for many months However for the highest data accuracy YSI recommends verifying the instrument s calibration on a daily basis To verify the instrument s calibration place the sensor in a water saturated air environment BOD bottle with 40 ml of water and check to see that the DO m...

Страница 7: ...section of this manual for instructions on how to determine if the sensor cap needs replaced The replacement OBOD sensor cap is shipped in a humidified container and the package should not be opened until the cap is needed for use Once the sensor cap has been installed on the OBOD probe as described in this section it is important to keep the sensor in a 100 humid environment Therefore the OBOD se...

Страница 8: ...ature portion of the sensor free of build up Other than that the sensor requires no maintenance A soft bristle brush can be used to scrub the temperature sensor if needed While cleaning be sure to only scrub the temperature sensor and not the sensor cap The sensor cap will be damaged if cleaned with anything abrasive Sensor Storage Short term Storage When the OBOD is not in use it must be stored i...

Страница 9: ... carefully remove the o ring and contact YSI Technical Support to obtain a new o ring and or gasket Avoid using tools to remove the o ring as damage to the sealing surfaces could result Before reinstalling the sensor cap make sure that the cavity is completely clean and dry before proceeding with the installation If water is found dry the cavity with lens tissue Finally clean and dry the clear opt...

Страница 10: ...ms to Australian and New Zealand EMC requirements for C Tick and RCM marks It conforms to unintentional radiator requirements under FCC Part 15 and ICES 003 Class B Manufacturer YSI Incorporated 1725 Brannum Lane Yellow Springs OH 45387 USA Product Name Professional ODO Water Quality Instrument and Professional OBOD probe Model Numbers Instrument Accessory ProODO 626281 ProComm Saddle manufactured...

Страница 11: ...8 technical support Email environmental ysi com Mail YSI Incorporated 1725 Brannum Lane Yellow Springs OH 45387 USA Internet ysi com When placing an order please have the following available 1 YSI account number if available 2 Name and phone number 3 Purchase Order or Credit Card 4 Model Number or brief description 5 Billing and shipping addresses 6 Quantity Service Information YSI has authorized ...

Страница 12: ...106 87 25 73 653 6 87 2 871 4115 1254 86 25 43 646 86 1 861 4430 1350 85 25 13 638 4 85 1 851 4747 1447 84 24 84 630 8 84 1 841 5067 1544 83 24 54 623 2 83 1 831 5391 1643 82 24 24 615 6 82 1 821 5717 1743 81 23 94 608 81 06 811 6047 1843 80 23 64 600 4 80 05 800 6381 1945 79 23 34 592 8 79 03 790 6717 2047 78 23 04 585 2 78 02 780 7058 2151 77 22 74 577 6 77 01 770 7401 2256 76 22 44 570 0 75 99 ...

Страница 13: ...Item 626471 Rev C Drawing 626471 April 2013 2013 YSI Incorporated ...

Отзывы: