Xylem YSI IDS ProOBOD Скачать руководство пользователя страница 7

6

7

UpDating inStrUment anD prOBe Firmware

The instrument and probe firmware can be updated via ysi.com. There 
you will find the new firmware files and instructions on how to update 
the instrument and/or probe. 

SenSOr maintenance - DiSSOlveD Oxygen

CLEANING ThE OBOD SENSOR CAP

The sensor cap should be kept clean since some types of fouling may 
consume or produce oxygen which could affect the dissolved oxygen 
measurements. To clean the sensor cap, gently wipe away any fouling 
with a lens cleaning tissue that has been moistened with water. Do not 
use a coarse towel or cloth or organic solvents to clean the sensor cap. 
Using a coarse towel or organic solvent to clean the sensor cap may 
cause permanent damage to the cap. For example, alcohol will dissolve 
the outer paint layer and other organic solvents will likely dissolve the 
dye in the cap.  If the sensor cap is damaged, it must be replaced.  

OBOD SENSOR CAP REPLACEMENT 

When replacing a sensor cap, the Sensor Cap Coefficients must be 
manually updated in the instrument. The instruction sheet shipped 
with the replacement OBOD sensor cap includes the coefficients.  See 
the instrument manual for instructions on updating the Sensor Cap 
Coefficients.

The frequency of sensor cap replacement is dependent on use. For 
example, when the probe is powered on for approximately 4 hours 
per day, 5 days a week, the sensor cap will need replaced about once 
per year.  If the probe is powered on more than this, the sensor cap will 
need replaced more often. 

To extend the sensor cap’s usable life, turn the instrument off when 
not in use. For example, turn the instrument off over night, but not 
between readings. 

The sensor cap will also need replaced if it is damaged or cracked. See 
the 

Troubleshooting 

section of this manual for instructions on how to 

determine if the sensor cap needs replaced. 

The replacement OBOD sensor cap is shipped in a humidified container 
and the package should not be opened until the cap is needed for 
use. Once the sensor cap has been installed on the OBOD probe as 
described in this section, it is important to keep the sensor in a 100% 

humid environment. Therefore, the OBOD sensor should be stored 
in either a BOD bottle that contains a small amount of clean water 
(approximately 40 ml) or in a BOD bottle  that is filled with clean water 
so the sensor cap is immersed in water, see

 

Sensor Storage for more 

information.  If the sensor dries out, refer to the rehydration procedure 
in this manual for instructions on how to recondition the sensor cap. 

Refer to Figure 2 below when following the instructions for replacing 
the cap. 

 

Figure 2.

Caution:  Avoid touching the sensing end of the sensor 

cap during the following maintenance procedures. 

1.  Remove the stir paddle from the probe by pulling it straight out.   
2.  Remove the old sensor cap from the probe by grasping the 

probe body with one hand and then rotating the sensor cap 
counterclockwise until it is completely free. Do not use any tools 
for this procedure. 

3.  Carefully remove the o-ring and discard it. Do not use any tools to 

remove the o-ring.

4.  Clean the o-ring seat on the probe by first wiping off old o-ring 

lubricant with a dry lens cleaning tissue. Next, clean away any 
build-up with a water-moistened lens tissue and then dry with 
another lens tissue.  

5.  Locate the o-ring supplied in the Sensor Cap replacement kit and 

install it on the probe. Do not use any tools to replace the o-ring.  
Be careful not to touch the clear optical DO sensor window. Any 
fingerprints on the optical window will have to be cleaned off. After 
installing the o-ring, ensure it is clean. If necessary, wipe clean with 
a lens tissue.

O-ring

 

OBOD Sensor Cap, 

shown here removed 

from probe

 

Содержание YSI IDS ProOBOD

Страница 1: ...USER MANUAL Document 626471 ProOBOD IDS ProOBOD ...

Страница 2: ...Dissolved Oxygen 4 Temperature 5 Taking Measurements 5 Maintenance and Storage 5 Updating Instrument and Probe Firmware 6 Sensor Maintenance Dissolved Oxygen 6 Sensor Maintenance Temperature 8 Sensor Storage 9 Troubleshooting 9 Dissolved Oxygen Readings 9 Error Status messages 11 Probe Specifications 11 Accessories Parts List 12 Declaration of Conformity 13 Recycling 14 Contact Information 14 Orde...

Страница 3: ...ervice Center selected by YSI Repair or replacement will be made and the product returned transportation prepaid Repaired or replaced products are warranted for the balance of the original warranty period or at least 90 days from date of repair or replacement LIMITATION OF WARRANTY This Warranty does not apply to any YSI product damage or failure caused by 1 Failure to install operate or use the p...

Страница 4: ...nt It fits in standard USA BOD bottles standard Japanese BOD bottles and many other international BOD bottles A funnel adapter for fitting in standard UK BOD bottles and a switching power supply with 3 international outlet adapters are also included IDS ProOBOD YSI item number 626500 This probe works with a MultiLab IDS instrument It fits in standard USA BOD bottles standard Japanese BOD bottles a...

Страница 5: ... contact YSI Customer Service at 800 897 4151 1 937 767 7241 or environmental ysi com or contact the authorized YSI distributor where the instrument was purchased Remove the OBOD probe cable assembly from the shipping container and locate the instruction sheet that is included This instruction sheet is important because it contains the calibration coefficients for your sensor cap A new probe cable...

Страница 6: ...ift than traditional electrochemical sensors This increased stability means that the instrument may hold its calibration for many months However for the highest data accuracy YSI recommends verifying the instrument s calibration on a daily basis To verify the instrument s calibration place the sensor in a water saturated air environment BOD bottle with 40 ml of water and check to see that the DO m...

Страница 7: ...section of this manual for instructions on how to determine if the sensor cap needs replaced The replacement OBOD sensor cap is shipped in a humidified container and the package should not be opened until the cap is needed for use Once the sensor cap has been installed on the OBOD probe as described in this section it is important to keep the sensor in a 100 humid environment Therefore the OBOD se...

Страница 8: ...ature portion of the sensor free of build up Other than that the sensor requires no maintenance A soft bristle brush can be used to scrub the temperature sensor if needed While cleaning be sure to only scrub the temperature sensor and not the sensor cap The sensor cap will be damaged if cleaned with anything abrasive Sensor Storage Short term Storage When the OBOD is not in use it must be stored i...

Страница 9: ... carefully remove the o ring and contact YSI Technical Support to obtain a new o ring and or gasket Avoid using tools to remove the o ring as damage to the sealing surfaces could result Before reinstalling the sensor cap make sure that the cavity is completely clean and dry before proceeding with the installation If water is found dry the cavity with lens tissue Finally clean and dry the clear opt...

Страница 10: ...ms to Australian and New Zealand EMC requirements for C Tick and RCM marks It conforms to unintentional radiator requirements under FCC Part 15 and ICES 003 Class B Manufacturer YSI Incorporated 1725 Brannum Lane Yellow Springs OH 45387 USA Product Name Professional ODO Water Quality Instrument and Professional OBOD probe Model Numbers Instrument Accessory ProODO 626281 ProComm Saddle manufactured...

Страница 11: ...8 technical support Email environmental ysi com Mail YSI Incorporated 1725 Brannum Lane Yellow Springs OH 45387 USA Internet ysi com When placing an order please have the following available 1 YSI account number if available 2 Name and phone number 3 Purchase Order or Credit Card 4 Model Number or brief description 5 Billing and shipping addresses 6 Quantity Service Information YSI has authorized ...

Страница 12: ...106 87 25 73 653 6 87 2 871 4115 1254 86 25 43 646 86 1 861 4430 1350 85 25 13 638 4 85 1 851 4747 1447 84 24 84 630 8 84 1 841 5067 1544 83 24 54 623 2 83 1 831 5391 1643 82 24 24 615 6 82 1 821 5717 1743 81 23 94 608 81 06 811 6047 1843 80 23 64 600 4 80 05 800 6381 1945 79 23 34 592 8 79 03 790 6717 2047 78 23 04 585 2 78 02 780 7058 2151 77 22 74 577 6 77 01 770 7401 2256 76 22 44 570 0 75 99 ...

Страница 13: ...Item 626471 Rev C Drawing 626471 April 2013 2013 YSI Incorporated ...

Отзывы: